Maselli
SA series - Grapes Analyzer - Maselli Misure
Automated grape must analyzer measuring alcohol content, pH, acidity, and color in under 90 seconds per cycle.
The SA Series Grapes Analyzer delivers rapid, multi-parameter analysis of incoming grape must at harvest reception. In a single automated cycle, it simultaneously measures alcohol concentration (0–18% vol.), pH (2–14), total acidity (0–30 g/l tartaric acid), and color tonality with high precision and temperature compensation. Processing 500 cc samples in 70–90 seconds, the analyzer reduces manual testing bottlenecks and ensures consistent quality data for every batch, enabling faster winemaking decisions at the point of delivery.
Specifications
- Application:
- Measure and analysis, in cycles with automatic loading, of a sample of grape must at the time of reception of grapes
- Limits/measurement scale:
- 0…18% vol. calculated conventionally on the basis of 16.83 g/l per 1% in volume of alcohol
- Accuracy:
- ±0.1% vol. maximum tolerated error is applied to indications which are not rounded off
- Concentration measurement:
- For instruments marketed in countries where type analysis is not envisaged
- PQ/Tonality measurement
- Measurement limits: PQ: 90…250 unità PQ Tonality: 0…2.5 Accuracy: PQ: 1 PQ unit Tonality: ±0.03
- Notes:
- The measurement is to be considered that on a turbid product (i.e. must)
- pH measurement
- Measurement limits: 2…14 pH Accuracy: ±0.05 pH unit
- Total acidity analysis
- Measurement scale span: 0…30 g/l of tartaric acid Accuracy: 0.10 g/l of tartaric acid
- Product temperature:
- 5…40 °C with automatic compensation of the temperature measured with Temperature Sensor Pt1000 in INOX 304, Class “A” in accordance with IEC751
- Quantity of analyzed sample:
- 500 cc for concentration measurement, 50 cc for analysis of pH and Total Acidity. In the version SA01/13, total quantity = ~630 cc
- Quantity of sample transferred:
- ~1500 cc/analysis cycle
- Capacity of transfer tank:
- ~5 litres; minimum 2 litres per cycle
- Cycle duration:
- Variable in accordance with time required for collection from trailer by the operator and programmed print parameters. Minimum 70 s, mean 90 s.
- Consumables:
- NaOH 2% Consumption ~5 ml/Analysis; pH4 and pH7 calibration solutions, Consumption ~25 cc. each/24 h